Z-0400-2024: BD Insyte Autoguard BC Shielded IV Catheter with Blood Control Technology 20GA X 1.00in, REF: 382533

Recall record Z-0400-2024 for Becton Dickinson Infusion Therapy Systems Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0400-2024
Company in FDA record
Becton Dickinson Infusion Therapy Systems Inc.
FDA classification date
Nov 24, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Over-the-needle, intravascular catheters may have needles that do not retract properly into the barrel, which could result in contaminated needlestick injury, blood leaking from the insertion site or catheter, and repeat catheter insertion attempts, due to catheter being inadvertently removed from the insertion site.

Source facts

Action
On 10/4/23, recall notices were emailed to customers and distributors who were asked to do the following: 1) Destroy all unused product subject to the recall following your institution's process for destruction. 2) For indwelling catheters, no further action is needed. 3) Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. 4) Complete and return the customer response form via email to BDRC48@bd.com Customers with recall questions, product complaints, or technical questions can call 1-844-823-5433, say recall when prompted, Mon-Fri 8:00am to 5:00PM CT, or email productcomplaints@bd.com In addition, distributors are asked to identify all customers within their distribution network that purchased affected product. Then distributors should provide a copy of the customer letter to these customers and advise them of the recall.
Initiated
2023-10-04
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution in the states of CA, NC, MI, WA, CO, FL, NJ, NM, IL, KY, TN, OH, TX, IN, LA, IA, MS, MO, NE, MD, NY, PA, GA, VA, ME, SC, RI, MA, PR.
Recall number
Z-0400-2024
Recall status
Open, Classified
CDRH record ID
203823
FDA classified
2023-11-24
Recall event number
93276

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗