The firm recalled the Freedom Cordless LED Headlight because some units were assembled with an incorrect component in the
control board electronics that may cause some Freedom Cordless LED Headlight units to become warm to the touch; however, no skin damage will occur.
Source facts
Action
Orascoptic sent an "URGENT: MEDICAL DEVICE RECALL" letter dated November 17, 2011. The letter identified the product, problem, and actions to be taken by the customers. An Acknowledgement Form was included for customers to complete and return to the firm. Contact Orascoptic Customer Care at 608-831-2555 for assistance regarding this notice.
Initiated
2011-11-17
Root cause
Labeling mix-ups
Terminated
2012-11-20
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada, Australia, and Europe.
Recall number
Z-0402-2013
Recall status
Terminated
CDRH record ID
114097
FDA classified
2012-11-20
Recall event number
63512