Reason stated in the source
An unrecoverable error may occur where a component will fail to complete the reboot cycle, become non-functional and must be replaced.
Source facts
- Action
- The firm disseminated a Medical Device Correction Notice to its consignees on 10/21/2021 by mail. Customer: Please share the information with potential users in your organization. Complete the attached response form and return to HillromCARY007@sedgwick.com within one month of receipt of this notice to confirm you have received this communication and understand the potential issue. In the event, this issue should occur please contact Hillrom Technical Support at 1-800-445-3720, option 3 or hrccwssupport@hillrom.com for service and replacement. Hillrom has remediated this issue as part of the next software upgrade to version 4.0.300. Please contact Hillrom to discuss software upgrade. Distributor: Please share this information with your nurse call customer base. Complete the attached response form and return to HillromCARY007@sedgwick.com within one month of receipt of this notice to confirm you have received this communication, understand the potential issue and will take action to inform your customer base.
- Initiated
- 2021-10-14
- Root cause
- Under Investigation by firm
- Terminated
- 2024-08-16
- Distribution
- Worldwide distribution - US Nationwide.
- Recall number
- Z-0402-2022
- Recall status
- Terminated
- CDRH record ID
- 189999
- FDA record created
- 2021-12-20
- Recall event number
- 88909