Open original FDA record ↗Recall record
- FDA identifier
- Z-0403-2013
- Company in FDA record
- Spacelabs Healthcare, Llc
- FDA classification date
- Nov 20, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Spacelabs Healthcare has learned through several reports that the input circuit may fail and the monitor will go on battery mode. The monitor will function normally until the batteries become depleted and then the monitor will shut down. In addition the monitor may fail to turn on due to an internal short.
Source facts
- Action
- Spacelabs Healthcare sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated October 19, 2012. The letter identified the product, problem, and actions to be taken by the customers. Contact Spacelabs Healthcare at 1-800522-7025 option 2 for questions regarding this notice.
- Initiated
- 2012-10-19
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-05-22
- Distribution
- Worldwide Distribution.On 4/16/13 the firm advised FDA that they added serial number 1390-000055 and 1390-000913. The two additional serial numbers were added because they were provided as replacements to a French customer whose devices had been returned to support the firm's investigation. The customer had already received the Urgent Medical Device correction, customer letter and required Spacelabs to provide these devices with the understanding that they would require the field corrective action
- Recall number
- Z-0403-2013
- Recall status
- Terminated
- CDRH record ID
- 114077
- FDA classified
- 2012-11-20
- Recall event number
- 63504