Skip to content

Public FDA record

Z-0403-2022: Hemashield Gold Knitted Bifurcated, 18 mm x 9 mm; reference number M002020851890

Recall record Z-0403-2022 for Maquet Cardiovascular, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0403-2022
Company in FDA record
Maquet Cardiovascular, LLC
FDA record creation date
Dec 20, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a probable mislabeling of product. A Hemashield Platinum Woven Double Velour Bifurcated Vasuclar Graft, SN 1259735369, 24 mm x 12 mm, REF Number M00202166241P0 is likely packaged inside of the carton instead of a Hemashield Gold Knitted Microvel Double Velour Bifurcated Vascular Graft; SN 1260004700; 18 mm x 9 mm; REF Number M002020851890 or a Hemashield Gold Knitted Microvel Double Velour Bifurcated Vascular Graft; SN 1259882880; 18 mm x 9 mm; Ref Number M002020851890.

Source facts

Action
And URGENT MEDICAL DEVICE REMOVAL notification letter dated 12/1/21 was sent to customers. Actions to be taken: " If you have already implanted one of the above-mentioned devices, kindly notify us as soon as possible by contacting your Getinge representative or Customer Service (at 1-888-943-8872, options 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST. You will be contacted shortly by our Medical Affairs Officer in order to discuss the appropriate follow up. " Please examine your inventory immediately to determine if you one of the Hemashield Gold Knitted Bifurcated Vascular Graft with a Serial Number matching those listed on page 1 of this letter. " Should you have un-used affected product, you are eligible for credit or a replacement at no cost to your facility upon receipt of Response Form (see page 4). " To get your free replacement Graft we need you to provide a ship to contact and your acknowledgment on page 4. " Please contact Getinge Customer Service at 1-888-943-8872, (press option 2) Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone), to request shipping instructions to return any affected product. " Pack the product to be returned with the appropriate return documents. " Whether you have affected product or not, please complete and sign the attached URGENT MEDICAL DEVICE REMOVAL RESPONSE FORM (page 4) to acknowledge this recall and return the completed form to Getinge by e-mailing a scanned copy to Hemashield.Intervascular@getinge.com or by FAX to 877-258-1332. " Please forward this information to all current and potential Hemashield Gold Knitted Microvel double velour vascular graft users within your hopsital / facility. " If you are a distrubutor who has shipped any affected product to customers, please forward this document to their attention for appropriate action. This voluntary recall only affects the product with the serial number and lot number listed on page 1; no other products
Initiated
2021-12-01
Root cause
Labeling Change Control
Terminated
2024-02-28
Distribution
US Nationwide distribution in the states of IL and NC.
Recall number
Z-0403-2022
Recall status
Terminated
CDRH record ID
190573
FDA record created
2021-12-20
Recall event number
89119

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗