Z-0404-2018: daVinci X EndoWrist(R) Stapler 45, REF 470298 Product Usage: The Intuitive Surgical EndoWrist Stapler 45

Recall record Z-0404-2018 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0404-2018
Company in FDA record
Intuitive Surgical, Inc.
FDA record creation date
Jan 23, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Specific da Vinci Xi EndoWrist Stapler 45 instruments are potentially impacted by a variation in the manufacturing process. In affected instruments, there is a possibility that the Stapler Release Kit (SRK) Wrench interface feature accessed through release hole 2 is manufactured to an incorrect dimension, which may prevent the SRK Wrench from being able to manually unclamp the instrument.

Source facts

Action
The firm initiated their recall by email on 09/19/2017 and followed with a letter delivered by FedEx on 09/20/2017. The notice requested the following actions: "Please take the following actions to ensure all affected personnel are fully informed of this Field Safety Notice. Forward this letter to your Risk Manager, OR Director, Purchasing, Biomedical Engineering staff, and other members of your medical staff who are involved with da Vinci procedures. 1.Using the lot numbers listed above, please locate and return all affected Staplers at your site via the standard RMA process. a. Credit will be provided for remaining uses on affected instruments returned to Intuitive Surgical. 2.Please log into the da Vinci Online Community field action resource to read and/or complete any requested actions related to this issue. a. https://www.davincisurgerycommunity.com/ 3.If you cannot access the da Vinci Online Community field action resource, please complete the attached Acknowledgement Form and return it via fax or email to Intuitive Surgical per the instructions contained in the Acknowledgement Form. 4.Please retain a copy of this letter and a copy of the Acknowledgement Form for your files."
Initiated
2017-09-19
Root cause
Process control
Terminated
2024-01-30
Distribution
US Nationwide Distribution
Recall number
Z-0404-2018
Recall status
Terminated
CDRH record ID
160837
FDA record created
2018-01-23
Recall event number
78908

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗