Skip to content

Public FDA record

Z-0405-2020: Aequalis Ascend Flex Humeral Stem: Part Number UDI DWF601A 3700386944475 DWF601B 3700386944482 DWF601C 37

Recall record Z-0405-2020 for Tornier, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0405-2020
Company in FDA record
Tornier, Inc
FDA record creation date
Nov 14, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During shipping, it is possible for the Aequalis Ascend Flex humeral stem to become unseated or otherwise move within its internal packaging restraints. Movement within the packaging may cause the product to rub against the PETG blister shipping container causing the transfer of PETG material from the blister container to the stem.

Source facts

Action
The firm notified their consignees by letter on 10/18/2019. The notices explained the issue and requested the following: Please undertake the following urgent actions when using Aequalis Ascend Flex humeral stems: " Ensure backup stem product is available in the operating room or is easily accessible prior to beginning surgery. " Visually examine the stem after removing the stem from its packaging. " If white film or particles are seen on the stem as pictured in Attachment A, immediately remove the implant and anything it touches from the sterile field. Replace the stem with a backup stem that lacks visible white film or particles. " If no packaging material is visible on the stem, you may proceed with the implantation procedure. " Report the discovery of any stems with packaging material to Wright Medical. " Contact customer service to facilitate the return and replacement of any affected product o Customer Service Customer.service@Wright.com, or 1-888-494-7950 o Return product to Wright Medical, 10801 Nesbitt Ave S, Bloomington, MN 55437 o Please identify that this product and issue is related to Field Action FA-WMG-2019-005 Please contact the following for information: Customer service Customer.service@Wright.com (replacements, product support) Field Action Team FieldAction@Wright.com (more information on this notice) Complaints Team Complaints-US@Wright.com (If product is found to be non-conforming)
Initiated
2019-10-18
Root cause
Package design/selection
Terminated
2021-04-09
Distribution
US Nationwide distribution.
Recall number
Z-0405-2020
Recall status
Terminated
CDRH record ID
177034
FDA record created
2019-11-14
Recall event number
84033

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.