Z-0405-2024: with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093, 006653, an accessory to

Recall record Z-0405-2024 for Breas Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0405-2024
Company in FDA record
Breas Medical, Inc.
FDA classification date
Nov 27, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Produced with unintended open slits on the side of the mouthpiece body, may result in insufficient ventilation in spontaneous breathing patients using MPV-VCV and MPV-PCV modes of ventilation

Source facts

Action
Breas Medical contacted customers via telephone followed by email on 10/13/23. Letter (Breas reference: SCAR-330) states reason for recall, health risk and action to take: Breas Medical recommends that all those potentially affected by this notice inspect existing inventory and identify mouthpieces which have been set up with the Vivo 45 LS for patient use. All defective mouthpieces shall be discarded. Breas Medical will provide replacement NIV Angled Mouthpieces, article number 008234, for all defective inventory and for devices in patient possession, where applicable. To obtain replacement NIV Angled Mouthpieces, we are asking customers to identify mouthpieces delivered from Breas since April 4, 2023 and complete the response form. We kindly ask you to acknowledge receipt of this Field Safety Notice together with your choice to remedy the issue using attached reply form NIV Angled Mouthpiece Field Safety Notice Distributor/Customer Reply Form. Contact Information Email: fsn@breas.com Helpline: +1 (855) 436-8724 Postal Address 16 Esquire Road, N. Billerica, MA 01862 Web Portal: breas.com
Initiated
2023-10-13
Root cause
Process control
Distribution
Nationwide Foreign: Canada
Recall number
Z-0405-2024
Recall status
Open, Classified
CDRH record ID
204016
FDA classified
2023-11-27
Recall event number
93331

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗