Skip to content

Public FDA record

Z-0406-06: BBL GasPak Pouch System and BBL GasPak CampyPouch System, Catalog numbers 260651 and 260656, packed in

Recall record Z-0406-06 for Becton Dickinson & Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0406-06
Company in FDA record
Becton Dickinson & Co.
FDA classification date
Jan 18, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Microbiological based test system may be contaminated with bacteria and cause false clinical test results in patient samples.

Source facts

Action
The recalling firm notified distributors by letter on 11/23/05 and advised that identified lots of the devices contain a component contaminated with the bacteria Ralstonia picketti. Distributors were instructed to discontinue distribution and discard. Distributors were requested to provide consignee lists to the recalling firm for direct notification to consignees. Distributors were notified that the recalled products have been discontinued. Customers are asked to discard Liquid Activating Reagent Packets and return the response form enclosed for replacement of the packets or alternatively use existing packets by adding user supplied de-ionized or sterile water into the reagent channel of the pouch.
Initiated
2005-11-23
Root cause
Other
Terminated
2006-02-01
Distribution
Product was distributed to direct accounts and distributors nationwide and to international affiliated distributors.
Recall number
Z-0406-06
Recall status
Terminated
CDRH record ID
43191
FDA classified
2006-01-18
Recall event number
34145

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗