Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
FDA record creation date
Nov 8, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
Vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download.
Source facts
Action
The firm began disseminating the Field Corrective Action notices by hand delivery on 10/02/2018 and followed with the posting of Medtronic Security Bulletin and the Urgent Medical Device Correction customer letter online on 10/11/2018.
The notice explained vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download and beginning October 11th, 2018, Medtronic will be disabling the SDN for programmer updates and will rely solely on the USB update method.
Initiated
2018-10-02
Root cause
Software design
Distribution
Worldwide
Recall number
Z-0406-2019
Recall status
Open, Classified
CDRH record ID
168165
FDA record created
2018-11-08
Recall event number
81300