Z-0407-2018: PURE OXYGEN CONCENTRATOR, CHAD Drive, Models CH5000 and CH5000S

Recall record Z-0407-2018 for Inovo, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0407-2018
Company in FDA record
Inovo, Inc
FDA record creation date
Jan 24, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Post-market surveillance for Pure Stationary Oxygen Concentrators model CHSOOO and CHSOOOS demonstrate an aggregate complaint/service rate greater than 35% for low oxygen purity (output oxygen below the published specification).

Source facts

Action
The firm initiated their recall by letter on 10/09/2017. The letter stated the following: "Actions to be Taken: Based on our distribution records, your firm has received one or more Pure Stationary Oxygen Concentrators. We respectfully request that you take the following actions: 1. Contact all patients and customers that currently have Pure Stationary Oxygen Concentrators in their homes or facilities and inform them of this recall. Retrieve the oxygen concentrators from their homes or facilities. 2. Destroy all Pure Stationary Oxygen Concentrator(s) in accordance with the following steps: a. Assure that the unit is not plugged into an electrical outlet. b. Cut the power cord of each oxygen concentrator. c. Take a digital photo of each oxygen concentrator that depicts both the power cord and the serial number of the depicted unit. Note that the serial number is located on a label on the back of the unit. d. Dispose of the oxygen concentrator in accordance with local ordinances. e. Record the serial number of each unit that is destroyed, sign and date the enclosed Recall Return Response Card. 3. Email or fax the completed response form and photos documenting destruction to: Email: lnovo4209@stericycle.com Fax: 1-877-497-2559 4. Sales of the Pure Stationary Oxygen Concentrator have been discontinued. Pure Stationary Oxygen Concentrators which have been destroyed will be replaced with a DeVilbiss 525DS Oxygen Concentrator in order to minimize any interruption in patient's supplemental oxygen therapy. 5. Pure Concentrators that have been returned to lnovo for repair will not be returned and will be replaced with a DeVilbiss 525DS Oxygen Concentrator. Other Information: Should you require any additional information about this recall action, please call 1-888-679-5132, Monday-Friday 8am-Spm Eastern."
Initiated
2017-10-09
Root cause
Under Investigation by firm
Terminated
2023-09-15
Distribution
worldwide
Recall number
Z-0407-2018
Recall status
Terminated
CDRH record ID
160803
FDA record created
2018-01-24
Recall event number
78883

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗