Z-0407-2024: Alinity m HCV AMP kit, List (REF) Numbers: a) 08N50-095 and b) 08N50-090. The Alinity m HCV assay is an

Recall record Z-0407-2024 for Abbott Molecular, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0407-2024
Company in FDA record
Abbott Molecular, Inc.
FDA classification date
Nov 29, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Abbott has received reports of an increase in reactive negative controls and false positive results in patient samples when using the Identified Lots of Alinity m HCV AMP Kit.

Source facts

Action
Abbott issued an Urgent Field Safety Notice to its consignees on 10/30/2023 via letter delivered using FedEx priority overnight express. The notice explained the problem, the potential impact/risk to patient, and requested the following actions: -discard all affected product -notify all those to which the products were transferred -follow internal laboratory procedures in the case of suspected incorrect results Credit or replacement will be issued for units discarded/destroyed. Abbott expanded the recall and notified consignees on 07/23/2024 via letter. The letter instructed consignees of an additional lot number included in the recall, but otherwise contained the same instructions.
Initiated
2023-10-30
Root cause
Other
Distribution
Worldwide
Recall number
Z-0407-2024
Recall status
Open, Classified
CDRH record ID
203730
FDA classified
2023-11-29
Recall event number
93230

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗