Z-0408-2019: CyberKnife M6, Part Number 054000-004 The CyberKnife M6 Systems are indicated for treatment planning an

Recall record Z-0408-2019 for Accuray Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0408-2019
Company in FDA record
Accuray Incorporated
FDA record creation date
Nov 8, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A robotics supplier notified Accuray of two manufacturing variations in the fastening of the in-line wrist and the casting of the wrist for the robot. The variation may result in premature failure of the component.

Source facts

Action
The firm, KUKA, issued a Field Safety Notice dated 9/13/2018 and disseminated it beginning on 09/26/2018 by mail, email, and hand delivery. The Field Safety Notice explained the issues and stated that all units would be inspected. Should a unit be faulty, the firm will arrange for remediation. If you have any technical questions, please feel free to get in touch with KUKA Account Manager or contact our hotline via phone 1-800-459-6691 or email: FocusCenter@Kuka.com.
Initiated
2018-09-26
Root cause
Process control
Terminated
2021-05-12
Distribution
worldwide distribution.
Recall number
Z-0408-2019
Recall status
Terminated
CDRH record ID
168358
FDA record created
2018-11-08
Recall event number
81352

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗