Z-0408-2024: Lugol s Iodine Solution 500ML, Part Number 624-71

Recall record Z-0408-2024 for EMD Millipore Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0408-2024
Company in FDA record
EMD Millipore Corporation
FDA classification date
Nov 29, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to the incorrect packaging utilized with lot 3124, there' s an improper seal between the cap and bottle. With an improper seal the Assay (current specification of 24.8 to 27.2 mL) is expected to continue to drop over time as iodine vapors continue to dissipate from the bottles.

Source facts

Action
A Product Recall notification letter dated 10/17/23 was sent to customers. Required/Recommended Action Please immediately examine your inventory to determine if you have any remaining impacted product and forward a copy of this letter to the personnel in your organization that utilize this product and/or need to be made aware of this issue. Please discard this material and request a replacement. Please complete attachment 1 form acknowledging receipt of this letter and return to the noted email address. In addition, if you may have further distributed this product, please identify your customers and notify them at once of this issue. Your notification to customers should include a copy of this notification. You may contact Technical or Customer Service to discuss an appropriate solution for your needs. If you would like more information about this FCA notification, please contact your local representative or use the link www.sigmaaldrich.com/offices for our website to obtain the phone number and/or email address of your local office.
Initiated
2023-10-17
Root cause
Under Investigation by firm
Distribution
US States: CA, NJ, WV
Recall number
Z-0408-2024
Recall status
Open, Classified
CDRH record ID
203929
FDA classified
2023-11-29
Recall event number
93318

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗