Open original FDA record ↗Recall record
- FDA identifier
- Z-0409-2013
- Company in FDA record
- KCI USA, Inc.
- FDA classification date
- Nov 20, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
The firm is initiating a voluntary medical device correction for the RotoRest Delta Therapy System head pack. KCI has received complaints alleging patients experiencing skin breakdown on the head and face due to the positioning of the head pack. In a few instances, the patient received injuries ranging from Stage III to Stage IV pressure ulcers.
Source facts
- Action
- KCI sent an Urgent - Voluntary Medical Device Correction letter dated June 27, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that KCI would contact them soon to coorinate a shipment of a replacement of the affected product. For any questions customers were instructed to call 1-800-275-4524, Selection Option 3 followed by Option 2. For questions regarding this recall call 1-800-275-4524.
- Initiated
- 2012-06-27
- Root cause
- Component design/selection
- Terminated
- 2013-04-22
- Distribution
- Nationwide Distribution including AL, CA, CO, OK and GA.
- Recall number
- Z-0409-2013
- Recall status
- Terminated
- CDRH record ID
- 111749
- FDA classified
- 2012-11-20
- Recall event number
- 62805