Completing Setup Without Program Move Complete Can Cause An Isocenter Error
Source facts
Action
Mevion issued an Important Safety Notice letter on October 31, 2016 to customers. The identified the affected product, problem, mitigation and advisory information. The letter informed customers that Mevion will release and install a software patch that removes the reliance on tracking isocenter changes for the correct reporting of the Couch position to the user displays and to the OIS. For further information mailto:srosenthal@mevion.com
Initiated
2016-10-31
Root cause
Software design
Terminated
2017-06-29
Distribution
US Nationwide in the states of FL, MO, NJ and OK
Recall number
Z-0409-2017
Recall status
Terminated
CDRH record ID
150527
FDA classified
2016-11-15
Recall event number
75496
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.