In rare circumstances, when surgeons use the software to plan the strut adjustment, an inverted image of the frame is generated by the software. The frame will appear inverted in the preview images on the Strut Settings tab.
Source facts
Action
The firm initiated their recall by email on 12/13/2017. Users are asked if they notice an inverted frame in the software to repeat the process for the same case and notify a Smith & Nephew representative. In addition, they are asked to double check plannings for strut adjustments and compare with the images printed as described above.
Initiated
2017-12-13
Root cause
Under Investigation by firm
Terminated
2020-10-05
Distribution
Worldwide
Recall number
Z-0409-2018
Recall status
Terminated
CDRH record ID
160710
FDA classified
2018-01-08
Recall event number
78840