Open original FDA record ↗Recall record
- FDA identifier
- Z-0410-2013
- Company in FDA record
- Zimmer, Inc.
- FDA classification date
- Nov 20, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Zimmer Inc. is initiating a removal of the Multipolar Bipolar Cup Liner 22 mm J.D. (item 00-5001-040-22) due to reports that the split poly ring was incorrectly assembled.
Source facts
- Action
- Zimmer sent an Urgent Medical Device Recall letter to all affected consignees during the week of September 3, 2012. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine inventory immediately, quarantine the recalled product and return remaining inventory. Customers were advised to complete Inventory Return Certification form attached.
- Initiated
- 2012-08-31
- Root cause
- Process design
- Terminated
- 2013-09-23
- Distribution
- Worldwide Distribution - USA nationwide including the states of: AK, CA, IL, FL, HI, MI, MO, NC, NJ, NY, OH, PA, TX and the countries of Germany, Brazil, Canada, Japan, Taiwan, Thailand, India and Korea.
- Recall number
- Z-0410-2013
- Recall status
- Terminated
- CDRH record ID
- 112642
- FDA classified
- 2012-11-20
- Recall event number
- 63122