Printer Support Module (PSM) may be subject to insufficient soldering, which may cause loosening of the wires and system to shut down. Affected devices need to have their Printer Support Modules replaced.
Source facts
Action
Firm notified impacted consignees by email on September 26, 2025. Consignees were informed that qualified service personnel will perform the correction of the device according to the standard service procedure.
Initiated
2025-09-26
Root cause
Process control
Distribution
Domestic distribution to CO, MI, MN, TN, TX. International distribution to Australia, Austria, Germany, Finland, India, Japan, Russia, China, Saudi Arabia.
Recall number
Z-0410-2026
Recall status
Open, Classified
CDRH record ID
216374
FDA classified
2025-10-29
Recall event number
97707