Software defect:2D projection of contours in Verity. The defect causes structures to appear off from their true position in the DRR and the radiograph with an error that increases with the distance of the structure from isocenter
Source facts
Action
Mevion issued an Important Safety Notice letter on October 14, 2016 to their customers. The letter identified the affected product, problem, containment, mitigation and actions taken. Customers were informed that Mevion will be updating software when it is released. For further Information, contact: srosenthal@mevion.com.
Initiated
2016-10-20
Root cause
Software design
Terminated
2017-05-15
Distribution
US Nationwide Distribution in the states of: FL, MO, NJ, OH, OK
Recall number
Z-0411-2017
Recall status
Terminated
CDRH record ID
150503
FDA classified
2016-11-15
Recall event number
75494
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.