Open original FDA record ↗Recall record
- FDA identifier
- Z-0411-2018
- Company in FDA record
- Baxter Healthcare Corporation
- FDA record creation date
- Jan 24, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Baxter has received customer complaints regarding leakage in the albumin circuit. The leakage was caused by an inadequate adhesive connection of the tubing to the Hansen connector of the MARS Tube Set which is part of the MARS Treatment Kit.
Source facts
- Action
- The firm initiated their recall on 12/21/2017 by letter. The recall notice stated the following actions to be taken by the customers: "1. Locate and remove all affected product from your facility. The product code and lot number can be found on the individual MARS Treatment Kit box. 2. Contact Baxter Healthcare Center for Service to arrange for return and credit. Baxter Healthcare Center for Service can be reached at 888-229-0001 between the hours of 7:00 am and 6:00 pm Central Time, Monday through Friday. Please have your Baxter 8-digit ship-to account number, product code, lot number, and quantity of product to be returned ready when calling. 3. If you purchased this product directly from Baxter, complete the enclosed Baxter customer reply form and return it to Baxter by faxing it to 224-270- 5457 or scanning and e-mailing it to fca@baxter.com, even if you do not have any inventory."
- Initiated
- 2017-12-21
- Root cause
- Nonconforming Material/Component
- Terminated
- 2020-08-04
- Distribution
- US, Germany, Sweden, Italy, Belgium, Portugal, Spain, France, Russia, Poland, Turkey, China, Thailand, Taiwan, Hong Kong, Malaysia, Korea, and Mexico.
- Recall number
- Z-0411-2018
- Recall status
- Terminated
- CDRH record ID
- 160660
- FDA record created
- 2018-01-24
- Recall event number
- 78835