Z-0412-2007: Thoratec Dual Drive Console a drive component of the Thoratec Ventricular Assist Device, Model Numbers, 1

Recall record Z-0412-2007 for Thoratec Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0412-2007
Company in FDA record
Thoratec Corp
FDA classification date
Feb 10, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The module power switch (at the back of console) may break while attempting to turn the Dual Drive Console OFF or ON.

Source facts

Action
The firm issued on 1/25/07 an URGENT: MEDICAL DEVICE CORRECTION-notice informing its consignees about the 'Power Switch Replacement Instructions' and arrangements for their service department replacement of the switches.
Initiated
2004-09-24
Root cause
Other
Terminated
2008-04-18
Distribution
Worldwide
Recall number
Z-0412-2007
Recall status
Terminated
CDRH record ID
48091
FDA classified
2007-02-10
Recall event number
36154

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗