Z-0412-2026: CellMek SPS Sample Preparation System, REF: C44603. Software versions 2.2.122.0 and prior, and Instructio

Recall record Z-0412-2026 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0412-2026
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Oct 30, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment.

Source facts

Action
Beckman Coulter notified consignees via email and letter on about 09/08/2025. For Issue 1 (Incorrect Dead Volumes), consignees were instructed to use the actual specimen dead volumes stated in Issue Description in the notification letter when processing samples in CellMek SPS with Sarstedt brand tubes. Beckman Coulter recommends sharing the content of the notification letter with laboratory and/or medical director regarding the need to review previous patient test results that were generated using the Sarstedt brand tubes. Beckman Coulter also recommends posting this letter on or near the affected instruments. Beckman Coulter will provide the correct CellMek SPS specimen dead volumes for Sarstedt brand tubes (Cassette Types B and E) in the next revision of the CellMek SPS Instructions for Use For Issue 2 (), consignees were instructed not to use Cassette Type E (15mm X 92mm Sarstedt brand tubes) to process samples on CellMek SPS. Beckman Coulter recommends sharing the content of the notification letter with laboratory and/or medical director regarding the need to review previous patient test results that were generated using the Sarstedt brand tubes. Beckman Coulter recommends posting the notification letter on or near the affected instruments. Beckman Coulter is actively investigating this issue and plans to address it in a future product release. Additionally, customers were instructed to notify personnel/customers if any affected units were transferred/distributed to another laboratory, and to complete and return the response form.
Initiated
2025-09-08
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, FL, IL, IN, MA, ME, MI, MN, MO, NY, OR, TN, UT, WA and the countries Australia, Austria, Belgium, Brazil, Canada, China, France, Germany, Greece, Hong Kong, Israel, Italy, Japan, Korea, Republic of, Mexico, Morocco, Netherlands, Poland, Portugal, Puerto Rico, South Africa, Spain, Switzerland, United Kingdom of Great Britain and Northern Ireland.
Recall number
Z-0412-2026
Recall status
Open, Classified
CDRH record ID
215868
FDA classified
2025-10-30
Recall event number
97531

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗