Reason stated in the source
Surgical gloves in hospital convenience kits may be degraded and tear during use.
Source facts
- Action
- Merit Medical Systems notified customers by phone, fax, or visit by sales representatives on 05/19/09. The notification advised of the convenience kit correction due to the recall of the Derma Prene Ultra Surgical Gloves, and that use of the product should cease immediately. Sales representatives were instructed to provide replacement gloves and "warning" stickers to be placed on identified kits in inventory. The warning stickers advise the customer to discard the surgical gloves in the kit and replace. Customers were requested to verify product disposition by signing a response form.
- Initiated
- 2009-05-19
- Root cause
- Nonconforming Material/Component
- Terminated
- 2009-12-03
- Distribution
- USA distribution: nine hospitals in Texas and Colorado.
- Recall number
- Z-0413-2010
- Recall status
- Terminated
- CDRH record ID
- 86098
- FDA classified
- 2009-11-23
- Recall event number
- 53619