Electrical failure of the bed due to fluid egress into the junction box.
Source facts
Action
In an October 21, 1998 e-mail the firm requested return of the product for a planned upgrade. The junction box cover was corrected at the same time of the upgrade.
Initiated
1998-10-21
Root cause
Other
Terminated
2003-01-28
Distribution
Nationwide
Recall number
Z-0414-03
Recall status
Terminated
CDRH record ID
25366
FDA classified
2003-01-09
Recall event number
25197