Open original FDA record ↗Recall record
- FDA identifier
- Z-0415-2007
- Company in FDA record
- Abbott Laboratories, Inc
- FDA classification date
- Feb 7, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The re-designed ARCHITECT Temperature Controller Board has been identified as being susceptible to electromagnetic interference (EMI) in the laboratory. This can cause a board reset condition, stopping temperature control function of the board. Situation does not stop analyzer operation and does not generate an error condition alerting operator when condition occurs.
Source facts
- Action
- Firm contacted consignees by telephone and fax with instructions to verify proper functioning of the Temperature Controller Board starting December 13, 2006.
- Initiated
- 2006-12-12
- Root cause
- Other
- Terminated
- 2007-09-24
- Distribution
- Worldwide, including USA, Argentina, Australia, Brazil, Canada, Chile, Columbia, Dominican Republic, Germany, Hong Kong, Jamaica, Japan, Korea, Mexico, New Zealand, Peru, Puerto Rico, Singapore, Taiwan, Thailand, Trinidad, Tobago, Uruguay, and Venezuela.
- Recall number
- Z-0415-2007
- Recall status
- Terminated
- CDRH record ID
- 49733
- FDA classified
- 2007-02-07
- Recall event number
- 36996