Open original FDA record ↗Recall record
- FDA identifier
- Z-0415-2016
- Company in FDA record
- MOOG Medical Devices Group
- FDA record creation date
- Dec 14, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for leaking between the purple ENFit connector and the white transitional stepped connector due to cracking or breaking of the purple connector.
Source facts
- Action
- MOOG Medical sent an URGENT Field Safety Notice, dated November 20, 2015, to its customer. The letter identified the affected device, provided a description of the problem, and discussed the risk to health. The customer was asked to contact all known distributors and inform them that MOOG Medical will cease production of all its enteral administration sets using ENFit connectors and transition back to the previous revision of the product codes that do not include the ENFit connector (INF0020, INF0500, INF1200, and GR1200). MOOG Medical will produce the previous revision until a solution to the problem is found. All necessary users should be made aware of the notice. Christopher Dodge, Manager, Regulatory Affairs, should be contacted with questions or for additional info at 801-264-1001, ext. 112 or cdodge@moog.com.
- Initiated
- 2015-11-20
- Root cause
- Unknown/Undetermined by firm
- Terminated
- 2016-08-22
- Distribution
- Distributed to the state of MN.
- Recall number
- Z-0415-2016
- Recall status
- Terminated
- CDRH record ID
- 141973
- FDA record created
- 2015-12-14
- Recall event number
- 72682