Skip to content

Public FDA record

Z-0415-2022: TruDi NAV Suction Instruments: 0-Degree, REF: TDNS000Z; 70-Degree, REF: TDNS070Z; and 90-Degree, REF: TDN

Recall record Z-0415-2022 for Acclarent, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0415-2022
Company in FDA record
Acclarent, Inc.
FDA record creation date
Dec 22, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Suction Instruments were incorrectly calibrated so they may not meet the system accuracy specification for visual verification of device location in the patient anatomy. If the surgeon does not detect the accuracy error, this could lead to an intraprocedural delay, a potential for reversible tissue damage or damage to the carotid artery, CSF leak, or damage to optic nerve.

Source facts

Action
On 11/12/21, recall notices were sent to OR Directors, Risk Managers, or Material Managers. Customers were asked to do the following: 1) Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). 2) Remove the product subject to this recall and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3) If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 4) Complete the Business Reply Form and return to acclarent8930@sedgwick.com 5) After the Business reply Form is received a return kit will be sent. 6) Keep this notice visibly posted for awareness until all product subject to this recall has been returned. Customers with additional questions are encouraged to call the recalling firm at 1-877-775-2789, 4:00am-5:00pm PST.
Initiated
2021-11-12
Root cause
Process control
Distribution
US: AZ, MI, IA, LA, IL, VA, SC, NY, IN, OR, SD, TX, PA, NE, FL, CT, MN, NV, CA, NJ, UT, AL, MA, WI, TN, WV, OH, NC, KY, HI¿, OK, GA, CO, MO, WY, MS, NH, HI, MD, ID
Recall number
Z-0415-2022
Recall status
Open, Classified
CDRH record ID
190691
FDA record created
2021-12-22
Recall event number
89170

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗