Z-0415-2026: Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-

Recall record Z-0415-2026 for Microgenics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0415-2026
Company in FDA record
Microgenics Corporation
FDA classification date
Oct 30, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm received complaints from customers reporting vial-to-vial variability when using the control with high sensitivity Troponin I assays. This defect may lead to as delay in patient results for hs Troponin I assays.

Source facts

Action
ThermoFisher Scientific issued an URGENT -MEDICAL DEVICE RECALL notice to its consignees on 10/02/2025 via email. The notice explained the issue, risk to health, and requested the following: "ACTIONS TO BE TAKEN BY A USER 1. If you are a user of the High sensitivity Troponin I Assay, determine if you have inventory of the affected lots of MAS Omni"CARDIO" control as listed in Table 1. As stated in the instructions for use, instrument values provided are specific to this lot of control only and are intended to assist the laboratory in establishing its own means and ranges. 2. You may continue to use controls that meet your laboratory s performance requirements. 3. If the variability of the material exceeds your laboratory s specification, dispose of the material in accordance with local, regional, and national regulations. After disposing of the material, contact TechService.mgc@thermofisher.com for a credit. 4. Customers using the affected product with standard/ regular Troponin I assays, can continue to use the product as long as it meets their laboratory s specification and continue monitoring performance of Troponin I in MAS Omni"CARDIO" control. 5. This recall notice needs to be passed to anyone within your organization or any organization where the potentially affected lots have been transferred. 6. Retain a copy of this letter for your laboratory records. 7. Please complete the Recall Response Form attached to this Notification within 10 days of the date of this letter and return to Thermo Fisher as instructed in the form to E-mail:MGCFSCA@thermofisher.com 8. If you have any additional questions regarding this Recall Notice, contact Thermo Fisher Scientific Technical Support at toll free 1-800-232-3342 (USA and Canada only) or 510-979-5000 (International), Option 2, then option 3 or by email: TechService.mgc@thermofisher.com." "ACTIONS TO BE TAKEN BY A DISTRIBUTOR 1. Please notify your customers of this Recall Notice using this recall notice and request t
Initiated
2025-10-01
Root cause
Nonconforming Material/Component
Distribution
Worldwide - US Nationwide distribution in the states of AL, OK, NH, NY, WI, IL, LA, AR, TX, UT, GA, ID and the countries of MEX, CAN, KOR, MYS, DEU, GBR, BRA, JAP, HKG, NZL, AUS.
Recall number
Z-0415-2026
Recall status
Open, Classified
CDRH record ID
216477
FDA classified
2025-10-30
Recall event number
97745

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗