The DNA-Probe has a target size of ~300 Kb region instead of the labeled ~190 Kb region.
Source facts
Action
Abbott Molecular sent letters dated October 2005 to their direct laboratory accounts on 10/28/05 via first class certified mail. The accounts were informed of the discrepancy between the actual size and region of coverage produced by the affected lots and that described in the product literature. The accounts were instructed to assess the impact of this discrepancy on the use of the product in their laboratory and to retain a copy of the letter with any of the affected product remaining in their inventory.
Initiated
2005-10-28
Root cause
Other
Terminated
2006-10-05
Distribution
Nationwide, Germany, Japan, Canada, Korea, India, Malaysia, Australia, Hong Kong, Chile & Singapore
Recall number
Z-0416-06
Recall status
Terminated
CDRH record ID
42506
FDA classified
2006-01-24
Recall event number
33825
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.