Skip to content

Public FDA record

Z-0416-2016: VentriClear II Ventricular Drainage Catheter Set. Allows external access and drainage of cerebrospinal f

Recall record Z-0416-2016 for Cook Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0416-2016
Company in FDA record
Cook Inc.
FDA classification date
Dec 14, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Cook Medical initiated a voluntary recall of specific lot numbers of the VentriClear II Ventricular Drainage Catheter Set due to the device being sterilized twice, which may result in the minocycline and rifampin catheter coating to be less than what is indicated on the labeling.

Source facts

Action
Cook Medical issued letter via certified mail on 11/19/2015, for their voluntary recall of the VentriClear II Ventricular Drainage Catheter Set due to being sterilized twice. The Distributor was instructed to forward the recall letter on to its customers, and quarantine any unused recalled devices, and return them to Cook Medical. Customers with questions may contact: Cook Medical Customer Relations 1-800-457-4500 or 1-812-339-2235. Monday through Friday between 7:30a.m. and 5:00p.m. Eastern Daylight Time or email at CustomerRelationsNA@cookmedical.com.
Initiated
2015-11-19
Root cause
Process control
Terminated
2017-02-22
Distribution
Distributed in Tennessee, Florida, Michigan, and Pennsylvania and in Saudi Arabia.
Recall number
Z-0416-2016
Recall status
Terminated
CDRH record ID
142021
FDA classified
2015-12-14
Recall event number
72718

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.