Z-0416-2022: Philips Allura Xper FD series with Software Version Number: 2.1.x- intended for Vascular, cardiovascular

Recall record Z-0416-2022 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0416-2022
Company in FDA record
Philips North America Llc
FDA record creation date
Dec 22, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Set screws securing the two gliding rods of the Extra Monitors (called 7th and 8th monitor) mounted on the Monitor Ceiling Suspension (MCS) can come loose, the Extra monitors might disengage from the Mounting Ceiling Suspension structure and fall down, may lead to patient, user or service engineer harm that may require medical intervention.

Source facts

Action
Philips issued Medical Device Correction letter to US consignees on 19-Nov-2021. Letters outside of the U.S.A. will be distributed through the Philips Markets Organization. Letter states reason for recall, health risk and action to take: Move the MCS with a 7th or 8th monitor only before or after a procedure. " Do not position the MCS above the patient/sterile field during the procedure. " Place this Field Safety Notice with the system documentation until your system has been fixed. " Circulate this notice to all users of this device so they are aware of the product issues. " Return the attached reply form to Philips to confirm that the users of the system have reviewed and understood this Field Safety Notice. A Philips representative will secure the set of screws of all affected systems. You will be contacted by your local Philips representative to schedule the service. If you need any further information, please contact your local Philips representative (reference to FC072200488).
Initiated
2021-11-19
Root cause
Other
Terminated
2024-02-07
Distribution
Worldwide distribution - US Nationwide distribution in the states of DC, GA, MA, TX, VA and the countries of Dominican Republic, INDIA, Japan, Poland, Spain, Taiwan.
Recall number
Z-0416-2022
Recall status
Terminated
CDRH record ID
190752
FDA record created
2021-12-22
Recall event number
89203

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗