Open original FDA record ↗Recall record
- FDA identifier
- Z-0417-2012
- Company in FDA record
- Biomet, Inc.
- FDA classification date
- Jan 11, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Biomet has initiated this action following an investigation which identified that Part 150367, (OSS Cemented 1M Stem 13x150) was incorrectly labeled as part 150368 (OSS Cemented 1M Stem 14x150).
Source facts
- Action
- Biomet, Inc. sent an "URGENT MEDICAL DEVICE RECALL NOTICE" dated November 1, 2011 to the one distributor with the affected product. The letter described the product, problem, and actions to be taken by the customer. The letter instructs the customer to immediately locate and remove the affected product from inventory. Additionally, the customer was asked to complete and return a Response Form via fax to 574-372-1683. Call 574-372-3983 for questions related to this notice.
- Initiated
- 2011-11-01
- Root cause
- Mixed-up of materials/components
- Terminated
- 2012-09-10
- Distribution
- Distributed only in The Netherlands.
- Recall number
- Z-0417-2012
- Recall status
- Terminated
- CDRH record ID
- 105423
- FDA classified
- 2012-01-11
- Recall event number
- 60447