- FDA identifier
- Z-0417-2013
- Company in FDA record
- SpineFrontier, Inc.
- FDA classification date
- Nov 21, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
When using the optional S-LIFT (also known as S-LIF) Implant Guides without stops, there is a known risk of moving the Implant Guides, and possibly the implant, too far into the disc space creating the potential for injury to the patient. The instructions for Use for the S-Lift Implant Guides will be amended to add the following warning: "Care is to be taken to not over insert the Implant Guides
Source facts
- Action
- Spine Frontier issued a Correction Notice Letter August 22, 2012 to all affected customers to update the Instructions for Use. The letter identified the affected product, problem and actions to be taken. Customers were Instructed to remove the Instructions for Use they currently have and replace them with the update. The letter instructs customers to complete and return the Response Form to 978-232-3990.
- Initiated
- 2012-08-22
- Root cause
- Labeling design
- Terminated
- 2015-06-11
- Distribution
- US Nationwide including the states of: MA and TX.
- Recall number
- Z-0417-2013
- Recall status
- Terminated
- CDRH record ID
- 112475
- FDA classified
- 2012-11-21
- Recall event number
- 63075