Skip to content

Public FDA record

Z-0417-2022: Stealth Autoguide Tracker, Model: 28248, which is an instrument in the Stealth Autoguide Basic Instrument

Recall record Z-0417-2022 for Medtronic Navigation, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0417-2022
Company in FDA record
Medtronic Navigation, Inc.
FDA record creation date
Dec 22, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A tracker that holds and guides neurosurgical surgical instruments may be subject to separation/weld failure, where the tracker connects to the instrument tube. If a weld separation occurs and a navigation inaccuracy is experienced, it may result in prolonged procedure and tissue injury. Per 11/24/21 firm email: The Bundle Autoguide Reg Kit does not have a model number. The Bundle Autoguide Reg Kit is internal terminology and not visible to the customer and are used as sales and shipping configuration only. When this sales bundle is chosen by an internal Medtronic Sales Representative the devices included in the bundle are individually packaged and labeled and sent to the customer in a single shipment. The devices included in the bundle are ordered, packed, and shipped as individual medical devices with unique model numbers, which is the reason that the Autoguide Tracker (28248) and the Stealth Autoguide Basic Instrument Kit (9736188) are the only part numbers subject to this FCA.

Source facts

Action
On 11/09/21, recall notices were distributed to healthcare professionals. The firm requested the following actions: 1) Identify, segregate, and quarantine affected products within your inventory. 2) Complete and return the Customer Confirmation Form even if you do not have inventory. The Customer Confirmation Form must be returned to initiate the Return Good Authorization (RGA) process. The RGA will be provided to you upon Medtronic receipt and processing of the Customer Confirmation Form. 3) Upon customer receipt of the RGA, return the affected product as instructed in the RGA. 4) This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. 5) Maintain a copy of this notice in your records. Customers with questions are encouraged to call: 1-888-826-5603. The recalling firm is working with its suppliers to make inventory available as soon as possible at which point a replacement device will be provided at no additional cost. Until replacement product is available, navigated alternatives can be used.
Initiated
2021-11-09
Root cause
Process control
Distribution
US: FL, NE, CA, MN, TX, OH, IL, AZ, KS, KY, OR, OK, WA, NV, MO, IN, MI, UT, MS, VA, SC. OUS: Canada, Sweden, Italy, France, Germany, Spain, Czech Republic, United Kingdom, India, Nepal, Japan, Turkey, China
Recall number
Z-0417-2022
Recall status
Open, Classified
CDRH record ID
190429
FDA record created
2021-12-22
Recall event number
89074

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗