A software defect exists in the ICIP Release 0.02 scheduling function for order management such that if a user edits either the Schedule Time or the Stop Time for an existing order, the corresponding intervention(s) for the order may not be created.
Source facts
Action
Philips Medical Systems notified accounts by letter on October 21, 2009, of the release ICIP version D.03, which contains a software correction . A Philips representative will schedule the implementation of this upgrade. Please contact the Philips Customer Care Center with questions or concerns about this correction at 1-800-722-9377.
Initiated
2009-10-21
Root cause
Software design
Terminated
2016-08-22
Distribution
MD, RI, WV, and TX
Australia
Belgium
Denmark
France
Italy
Malaysia
Singapore
Spain
Switzerland
Taiwan
United Arab Emirates
United Kingdom
and
Ireland
Recall number
Z-0418-2010
Recall status
Terminated
CDRH record ID
85943
FDA classified
2009-11-24
Recall event number
53593