Open original FDA record ↗Recall record
- FDA identifier
- Z-0418-2015
- Company in FDA record
- Materialise USA LLC
- FDA classification date
- Nov 21, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The box of guides for a specific case arrived intact but contained two (2) femoral guides instead of one (1) femoral guide and one (1) tibial guide. The device did not meet specification: One tibia guide missing and one femur guide contained an edge which was +0.04 mm over specification.
Source facts
- Action
- Materialise USA was notified on October 16, 2014 by the Field Representative of the distributor (Zimmer) that the box of guides for case JTOMA-JL-0636-L arrived intact but contained two (2) femoral guides instead of one (1) femoral guide and one (1) tibial guide. A new set of tibial and femoral guides were manufactured and sent to Zimmer (headquarters distribution center) on October 17, 2014 before the surgery date on October 20, 2014. For questions regarding this recall call 734-662-5057.
- Initiated
- 2014-10-16
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-02-17
- Distribution
- Nationwide Distribution to MN only
- Recall number
- Z-0418-2015
- Recall status
- Terminated
- CDRH record ID
- 131198
- FDA classified
- 2014-11-21
- Recall event number
- 69632