Skip to content

Public FDA record

Z-0419-2007: Cytomics FC 500 MPL with MXP software, Versions 2.0 & 2.1: Part #731419 KIT; FOC MXP 2.0 UPGRADE KIT; Par

Recall record Z-0419-2007 for Beckman Coulter Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0419-2007
Company in FDA record
Beckman Coulter Inc
FDA classification date
Feb 8, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Two workflow scenarios associated with renaming regions may produce incorrect results.

Source facts

Action
A Product Corrective Action (PCA) letter was sent on the week of August 29, 2006 to all accounts that have an FC500 or have ordered CXP or MXP software to inform them that a) incorrect statistics are generated when a polygonal region with vertical re-entrant vertices is renamed using the Region Properties dialog box. b)when a region is copied from one plot to another, and then the new region is renamed, the statistics are not updated with the new region name. (Letter was sent by US mail). Letter includes immediate user preventive action to remedy both issues. A response form is included for consignees to return to recalling firm.
Initiated
2006-08-29
Root cause
Other
Terminated
2012-04-19
Distribution
Nationwide & Canada
Recall number
Z-0419-2007
Recall status
Terminated
CDRH record ID
49917
FDA classified
2007-02-08
Recall event number
37083

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗