It was identified that the Mistral-Air Premium blankets may be capable of
conducting electricity during a defibrillation or cardioversion therapy at higher
energy levels (70 Joules or more) when the blanket is applied with the reflective
side toward the patient (upside down).
Source facts
Action
Stryker notified customers of the recall by sending Customer Notification and Business Reply Forms, dated July 21, 2017.
Initiated
2017-07-21
Root cause
Component design/selection
Terminated
2018-03-22
Distribution
U.S.
Recall number
Z-0419-2018
Recall status
Terminated
CDRH record ID
160806
FDA record created
2018-01-27
Recall event number
78886