Skip to content

Public FDA record

Z-0420-03: Pharmaseal Thoracentesis Tray with Catheter, catalog number 4341B; a sterile, single use, Rx, latex free

Recall record Z-0420-03 for Cardinal Health. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0420-03
Company in FDA record
Cardinal Health
FDA classification date
Jan 9, 2003
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Raw material used in the manufacture of catheters may contain a contaminant predisposing the catheter to become brittle, possibly breaking during use.

Source facts

Action
Recall letters dated 11/15/02 were sent via 2-day FedEx to the direct accounts. The accounts were informed of the potential for breakage of the catheters during use, requested to inspect their inventories for the affected lot numbers, and segregate any found for return to Allegiance for replacement. Distributors were requested to sub-recall the product from their accounts. The recall was extended to include the 21 additional lots by letter dated 4/8/04. The letters were sent via Airborne Express, 1-Day, on 4/8/04. The instructions were the same as the 11/15/02 letter, except that Allegiance has changed their name to Cardinal Health.
Initiated
2002-11-15
Root cause
Other
Terminated
2006-04-14
Distribution
Nationwide distribution, including Puerto Rico and Guam, and internationally to Saudi Arabia
Recall number
Z-0420-03
Recall status
Terminated
CDRH record ID
25087
FDA classified
2003-01-09
Recall event number
25054

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗