Open original FDA record ↗Recall record
- FDA identifier
- Z-0420-2015
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA classification date
- Nov 21, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Certain lots of the reagent cartridges show confirmed positive bias averaging 0.4% Hemoglobin A1c units and occasionally up to 1.0% HbA1c units for patient samples when compared to the National Glycohemoglobin Standardization Program (NGSP). QC samples may exhibit a similar bias.
Source facts
- Action
- Siemens sent an Urgent Medical Device Correction letter dated September 2014 was provided to all Dimension Vista HbA1c and Dimension HB1C customers who received the affected lots to notify them of the issue, the potential risk to health, and actions to be taken by customers. Customers were instructed to discontinue use of and discard the affected devices. Customers need to review the letter with their Medical Director. Siemens will replace any unused inventoryof the affected lots at no charge. Customers were asked to complete the attached form and fax it to 302-631-7795. Customers should retain a copy of the letter with their laboratory records, and forward the letter to those who may have received this product. Customers with questions should contact their Siemens Customer Care Center - Technical Solutions or their local Siemens technical support representative. For questions regarding this recall call 800-441-9250.
- Initiated
- 2014-09-30
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-08-10
- Distribution
- Worldwide Distribution: US (nationwide) and internationally to: Canada.
- Recall number
- Z-0420-2015
- Recall status
- Terminated
- CDRH record ID
- 130530
- FDA classified
- 2014-11-21
- Recall event number
- 69434