Z-0420-2020: EZ Glide Aortic Cannula, 24 Fr. (8.0mm) x 37.6 cm (14.8") 3/8" (9.5 mm) vented connector , RX Only, Steri

Recall record Z-0420-2020 for Edwards Lifesciences, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0420-2020
Company in FDA record
Edwards Lifesciences, LLC
FDA record creation date
Nov 18, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Their is a possibility that the cannula may separate from its connector, potentially causing a breach of circuit during cardiopulmonary bypass which could result in significant blood loss.

Source facts

Action
The firm plans to begin mailing a "urgent - Product Recall" notification on 10/22/2019 via FedEx Next Day Delivery. All recall notification letters should be mailed by 10/23/2019. The customer notification will request consignees take the following actions: 1. Review this field safety notice to understand the potential hazard, and return all unused inventory as instructed. 2. Complete and return the Product Reconciliation Form to Customer Service: a) Record the quantity of any affected EZ Glide aortic perfusion cannula in your possession, b) Segregate and quarantine affected product until returned, c) Determine the quantity of EZ Glide aortic perfusion cannula used by subtracting the quantity on hand from the quantity shipped to you, d) Contact Customer Service to arrange return of affected devices, and e) Return affected devices to Edwards with the Return Goods Authorization (RGA) provided. 3. Complete and return the attached Acknowledgement Form within five (5) business days of receiving this notice to Customer Service via fax at (800) 422-9329 or email to cs.usfaxes@edwards.com. 4. Distribute this notice within your organization or to any organization where the potentially affected devices have been transferred. If you have further distributed this product, notify your customers to the user level. Report any EZ Glide separation to Edwards Lifesciences. 5. The firm reccomends using an alternative device. However if another device is not available the firm provides the following instructions: Should you elect to use the EZ Glide device in your inventory, we suggest grasping each side of the bond area and applying a firm pull before use to confirm the bond is secure. We also recommend maintaining visibility of the device throughout the procedure. Potential mitigations in the event of cannula separation include ensuring a backup device or 3/8 x 3/8 connector is available. 6. Edwards has communicated this Field Safety Notice to ap
Initiated
2019-10-22
Root cause
Under Investigation by firm
Terminated
2022-06-29
Distribution
Worldwide distribution - US Nationwide distribution in the states of Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Washington DC, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, Tennessee, Texas, Utah, Virginia, Vermont, Washington, Washington DC, Wisconsin, West Virginia, Wyoming and countries of Austria, Australia, Belgium, Bahrain, Canada, Switzerland, Chile, China, Cyprus, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hungary, Canary Islands, Ireland, India, Iceland, Italy, Japan, Mauritius, Malaysia, Netherlands, Norway, Poland, Romania, Sweden, Singapore, Slovakia, Thailand, Turkey, United Arab Emirates, South Africa.
Recall number
Z-0420-2020
Recall status
Terminated
CDRH record ID
177110
FDA record created
2019-11-18
Recall event number
84065

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗