Z-0420-2026: TruDi Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-

Recall record Z-0420-2026 for Integra LifeSciences Corp. (NeuroSciences). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0420-2026
Company in FDA record
Integra LifeSciences Corp. (NeuroSciences)
FDA classification date
Nov 4, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was determined that when the navigation system is configured with specific revision combinations of the Multi Instrument Adapter together with the Patient Tracker the system may not meet its specified accuracy for visual verification of device location within the patient s anatomy.

Source facts

Action
On October 3, 2025 URGENT: VOLUNTARY MEDICAL DEVICE RECALL letters were sent to customers. Actions to be Taken by Customers 1. Examine the your system to determine if you have product on hand subject to this field recall (Rev 00 and Rev01). 2. If you have revision 00 and/or revision 01, please check the box on the enclosed Acknowledgement Form I do have affected product(s) and record the total quantity of the affected product that you have. 3. If you do not have affected product, check the box, I do not have affected product. 4. Complete the rest of the Acknowledgement Form and return. 5. Keep a copy of the form for your records. 6. Forward this notice to those who utilize the product so they are aware of this recall and can identify and quarantine any affected product that may remain in clinical areas. 7. When your form is received, and it is noted that you have Rev 00 or Rev 01 products, Customer Service will contact you and provide a Return Material Authorization (RMA) number and directions to return the affected product. You can request a replacement which you will receive depending on availability. 8. In an effort to prioritize replacement Adapters please note below (in the Medical Facilities Acknowledgement Form) if you have Rev 02 and Rev 03 to use for your upcoming planned surgeries.
Initiated
2025-10-03
Root cause
Device Design
Distribution
US Nationwide distribution.
Recall number
Z-0420-2026
Recall status
Open, Classified
CDRH record ID
216327
FDA classified
2025-11-04
Recall event number
97677

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗