Skip to content

Public FDA record

Z-0422-05: Ultraview Universal Clinical Workstation System Model 90385.

Recall record Z-0422-05 for Spacelabs Medical Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0422-05
Company in FDA record
Spacelabs Medical Incorporated
FDA classification date
Jan 25, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Weakness of the swivel base for the universal clinical workstation may result in the unit falling from its mounted position endangering users and patients.

Source facts

Action
All US customers were sent a Medical Device Recall letter on 1/7/2005 describing how to examine monitors for the defect and to discontinue use of any suspect unit s until examintaion by a Field Service Engineer . International customers will be notified via email or letter on 1/14/2005. Follow up contact by Customer Support and Global Support is planned for any customer not responding with a FAX-back form.
Initiated
2005-01-07
Root cause
Other
Terminated
2011-01-17
Distribution
The firm completed a survey of their customer records and determined a total of 12,306 units were installed from January of 1994 through December of 1997. 7,601 units were originally sold to 767 US accounts and 4,705 units were sold to 102 international accounts. The firm estimates approximately 3000 units worldwide are still in operation with the Revision M or earlier base.
Recall number
Z-0422-05
Recall status
Terminated
CDRH record ID
36615
FDA classified
2005-01-25
Recall event number
30797

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗