Acoustic output intensity may exceed maximum specified limits and the patient contact temperature may exceed 43 degree C (when the system is used with the S12-4 transducer in CW-mode)
Source facts
Action
On 11/2/06 the firm sent a letter dated October 31, 2006 to their customers. The letter explains the problem and states that new software is being released and will be available sometime in November. Customer service representatives will visit consignees to upgrade their units. The firm later determined they sent letters out to consignees that did not have affected units. Those consignees will be notified the week of 11/20/06 to disregard the previous letter.
Initiated
2006-11-02
Root cause
Other
Terminated
2009-07-14
Distribution
Worldwide
Recall number
Z-0422-2007
Recall status
Terminated
CDRH record ID
49224
FDA classified
2007-02-17
Recall event number
36722