Open original FDA record ↗Recall record
- FDA identifier
- Z-0422-2011
- Company in FDA record
- Davol, Inc., Sub. C. R. Bard, Inc.
- FDA classification date
- Nov 19, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Mislabeled: Hernia Patch may be a different size than on the product label
Source facts
- Action
- Davol notified accounts by an Urgent: Medical Device Recall letter on 10/15/10 via overnight delivery. The letter stated that the product maybe a different size than indicated on the product label. It also identified the affected product. Accounts are requested to examine their inventory, remove, and return recalled product. Customers are to notify any of their customers that they may have further distributed the affected product to by including a copy of the recall notice. Customers should contact Davol Customer Service 1-800-556-6275 for instructions on how to return the affected product. Customers are also requested to complete and fax back the attached fax sheet to 1-401-825-8753. If there are any additional questions, customers should contact Davol Customer Service Department at 1-800-556-6275 or C.R. Bard's Medical Service Support Department at 1-800-562-0027.
- Initiated
- 2010-10-15
- Root cause
- Packaging process control
- Terminated
- 2011-10-27
- Distribution
- Nationwide Distribution
- Recall number
- Z-0422-2011
- Recall status
- Terminated
- CDRH record ID
- 95172
- FDA classified
- 2010-11-19
- Recall event number
- 57010