Z-0422-2023: Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900)

Recall record Z-0422-2023 for Getinge Usa Sales Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0422-2023
Company in FDA record
Getinge Usa Sales Inc
FDA classification date
Dec 7, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event of a mains power failure. This could require the clinician to use manual emergency ventilation of another ventilator is not available and the backup battery is exhausted.

Source facts

Action
On 11/7/22 the firm notified customers via Urgent Medical Device Correction letters. Customers were instructed to immediately identify whether they had affected Flow-c or Flow-e anesthesia systems on hand. Customers who have affected devices may continue to use the device ensuring that a system checkout procedure is performed per the Operators Manual. The firm will replace the Power backup battery free of charge for customers with affected anesthesia systems. You can also arrange a visit by a Getinge trained or authorized service technician by contacting Getinge Customer Technical Support at (888) 9GETUSA / (888) 943 8872 (select option 1, then option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Initiated
2022-11-07
Root cause
Component design/selection
Distribution
US Nationwide distribution in the states of CO, FL, IN, MA, MO, and VA.
Recall number
Z-0422-2023
Recall status
Open, Classified
CDRH record ID
196696
FDA classified
2022-12-07
Recall event number
91088

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗