Z-0422-2024: RAPIDPOINT 500e Blood Gas System

Recall record Z-0422-2024 for Siemens Healthcare Diagnostics Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0422-2024
Company in FDA record
Siemens Healthcare Diagnostics Inc
FDA classification date
Dec 1, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There are potential drug interferences from Perhexiline Maleate or Atomoxetine Hydrochloride that may cause falsely elevated sodium results reported on the RAPIDPoint 500 and RAPIDPoint 500e Blood Gas Systems. The erroneous sodium levels with positive bias may lead to unrecognized hyponatremia and/or the inappropriate treatment of hypernatremia may lead to iatrogenic hyponatremia, resulted in a delay in patient diagnosis and optimal patient management.

Source facts

Action
The firm issued an Urgent Medical Device Correction notice on 9/26/2023 to customers asking that customers review the notice with the facility's Medical Director to determine the appropriate course of action for previously generated results, complete and return the Field Correction Effectiveness Check Form within 30 days, forward this notice to customers if your firm is a distributor, and retain the notification in your system's Operator's Guide and forward to those who may have received product. The Field Correction Effectiveness Check Form is to be returned via email to uscctsfcaecfax.team@siemens-healthineers.com or by fax to 312-275-7795. Questions are to be directed to your local Siemens Healthineers Technical Support Representative.
Initiated
2023-09-26
Root cause
Under Investigation by firm
Distribution
Domestic: Nationwide Distribution. Foreign: Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Congo, Republic, Costa Rica, Cote dIvoire, Croatia, Curacao, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Fiji, Finland, France, French, Polynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Lesotho, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Martinique, Mexico, Moldova, Montenegro, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestinian Territory, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom Great Britain, Uruguay, & Vietnam.
Recall number
Z-0422-2024
Recall status
Open, Classified
CDRH record ID
203741
FDA classified
2023-12-01
Recall event number
93240

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗