Skip to content

Public FDA record

Z-0423-2023: Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Prog

Recall record Z-0423-2023 for Insulet Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0423-2023
Company in FDA record
Insulet Corporation
FDA classification date
Dec 7, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan.

Source facts

Action
On November 14, 2022, the firm initiated their Press Release at https://investor.insulet.com/news-releases/news-release-details/insulet-issues-nationwide-voluntary-medical-device-correction and notified affected customers through Urgent Medical Device Correction letters. Customers were instructed to monitor their Omnipod 5 Controller for the following issues: - Charging port or charging cable appears melted, deformed, or discolored - Controller experiences extreme overheating (uncomfortable to hold) or emits an odor while charging If customers experience any of the issues above, they should disconnect the charger from the power outlet and do NOT charge the Omnipod 5 Controller. Promptly contact Insulet s dedicated Customer Care team specifically trained to help you at 1-800-641-2049 to request a replacement device. If you have a compatible Android smartphone (please see omnipod.com/compatibility), you may want to switch to using the Omnipod 5 App on your smartphone. Customers were provided with additional precautions to take to continue use of the device.
Initiated
2022-11-14
Root cause
Device Design
Distribution
Domestic distribution nationwide.
Recall number
Z-0423-2023
Recall status
Open, Classified
CDRH record ID
196863
FDA classified
2022-12-07
Recall event number
91127

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.