Z-0425-2021: Puritan Bennett 980 (PB980) Ventilator Series - Product Usage: for use on patient population sizes from N

Recall record Z-0425-2021 for Covidien Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0425-2021
Company in FDA record
Covidien Llc
FDA record creation date
Nov 6, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In-house lab observations and customer reports of overheating and/or thermal damage to the battery pack, create the risk of exposure to smoke or hazardous materials, which may result in burns and/or bronchospasm,

Source facts

Action
Medtronic issued Urgent Medical Device Correction letter to consignees via Federal Express and/or certified mail beginning October 16, 2020. The letter states reason for recall, health risk and action to take: Immediately notify all care environments in which the PB980 series ventilators are used about this letter. If the facility has distributed PB980 series ventilators to other persons or facilities, consignees are directed to promptly forward a copy of the letter to those recipients. Complete the response form and return it as directed to confirm receipt and understanding of the information. Adhere to the instructions described in the PB980 Series Ventilator Operators Manual and battery addendum for Li-Ion Battery Pack Handling, Storage, and Replacement. Promptly dispose of battery packs that are faulty or have reached their end-of-useful-life, according to local regulations. informs customers of the release of software update, R7.1, for all models of Puritan Bennett" 980 (PB980) series ventilators. Actions being taken by Medtronic Medtronic developed a software update, R 7.1. This software release is available for installation. For all warranty and service contract customers, Medtronic Service will load the R 7.1 software during your next scheduled or planned maintenance visit. For those customers outside of warranty or not on an active service agreement, when your response form is received Medtronics Service Team will contact you to set up a time to perform the update In addition, Medtronic made modifications to the PB980 Series Ventilator Operators Manual. The manual has been updated with additional information for users regarding the new software update. The revised manual can be accessed at: https://www.medtronic.com/covidien/en-us/support/product-manuals.html Questions, please contact our Technical Support Department at 1-800-255-6774
Initiated
2020-10-16
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Argentina, Bolivia, Brazil, Cayman Islands, Chile, Columbia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Jamaica, Mexico, Nicaragua, Panama, Peru, Uraguay, Honduras, Austria, Bahrain, Belgium, Bulgaria, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Jordan, Kuwait, Latvia, Lebanon, Lithuania, Netherlands, Norway, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Taiwan, Korea, Brunei Daruss., Cambodia, Indonesia, Malaysia, Singapore, Thailand, Vietnam, Japan.
Recall number
Z-0425-2021
Recall status
Open, Classified
CDRH record ID
184238
FDA record created
2020-11-06
Recall event number
86643

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗