Open original FDA record ↗Recall record
- FDA identifier
- Z-0425-2021
- Company in FDA record
- Covidien Llc
- FDA record creation date
- Nov 6, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
In-house lab observations and customer reports of overheating and/or thermal damage to the battery pack, create the risk of exposure to smoke or hazardous materials, which may result in burns and/or bronchospasm,
Source facts
- Action
- Medtronic issued Urgent Medical Device Correction letter to consignees via Federal Express and/or certified mail beginning October 16, 2020. The letter states reason for recall, health risk and action to take: Immediately notify all care environments in which the PB980 series ventilators are used about this letter. If the facility has distributed PB980 series ventilators to other persons or facilities, consignees are directed to promptly forward a copy of the letter to those recipients. Complete the response form and return it as directed to confirm receipt and understanding of the information. Adhere to the instructions described in the PB980 Series Ventilator Operators Manual and battery addendum for Li-Ion Battery Pack Handling, Storage, and Replacement. Promptly dispose of battery packs that are faulty or have reached their end-of-useful-life, according to local regulations. informs customers of the release of software update, R7.1, for all models of Puritan Bennett" 980 (PB980) series ventilators. Actions being taken by Medtronic Medtronic developed a software update, R 7.1. This software release is available for installation. For all warranty and service contract customers, Medtronic Service will load the R 7.1 software during your next scheduled or planned maintenance visit. For those customers outside of warranty or not on an active service agreement, when your response form is received Medtronics Service Team will contact you to set up a time to perform the update In addition, Medtronic made modifications to the PB980 Series Ventilator Operators Manual. The manual has been updated with additional information for users regarding the new software update. The revised manual can be accessed at: https://www.medtronic.com/covidien/en-us/support/product-manuals.html Questions, please contact our Technical Support Department at 1-800-255-6774
- Initiated
- 2020-10-16
- Root cause
- Device Design
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Argentina, Bolivia, Brazil, Cayman Islands, Chile, Columbia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Jamaica, Mexico, Nicaragua, Panama, Peru, Uraguay, Honduras, Austria, Bahrain, Belgium, Bulgaria, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Jordan, Kuwait, Latvia, Lebanon, Lithuania, Netherlands, Norway, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Taiwan, Korea, Brunei Daruss., Cambodia, Indonesia, Malaysia, Singapore, Thailand, Vietnam, Japan.
- Recall number
- Z-0425-2021
- Recall status
- Open, Classified
- CDRH record ID
- 184238
- FDA record created
- 2020-11-06
- Recall event number
- 86643